Product Specification
Brand | ANTAR TECHNOLOGY |
Condition | New |
Product Descriptions
Technical Parameter
Type |
AT-L-750 |
Voltage (V) |
220 (1~) |
Frequency (Hz) |
50/60 |
Power (KW) |
0.75 |
Rated current (A) |
4.1 |
Motor speed (r/min) |
2800 |
Maximum pressure (KPa) |
12 |
Maximum vacuum (KPa) |
-12 |
Number of dental chairs (pcs) |
1~2 |
Noise (db) |
62 |
Package size (mm) |
54*38*70 |
Package weight (kg) |
36 |
1.Motive power system adopt width voltage & double frequency ,motor with protection IP55 grade as driving, to accomplish turbine circle suction strength with lower nosie , higher pressure ,larger rate of flow
2.Filter system adopt advanced mechanical filter technology in the present water technology,can effective dislodge rubber mass and other solids
3.high efficiency & twice water and air separation system can prevent liquid & blood and bloody bulla of pipeline to suction vacuum pump.high foam compatibility. During high rate of flow, it still have stable separation efficiency. small obstruction , simple maintenance.
Semi-Dry Suction system
Advantage
1. Motive power system adopt width protection IP55 grade as driving, to accomplish turbine circle suction strength with lower nosie , higher pressure ,larger rate of flow
2. Filter system adopt advanced mechanical filter technology in the present water technology,can effective dislodge rubber mass and other solids
3. high efficiency & twice water and air separation system can prevent liquid & blood and bloody bulla of pipeline to suction vacuum pump.high foam compatibility. During high rate of flow, it still have stable separation efficiency. small obstruction , simple maintenance.
Struction and working theroy
1. Inside major structure--Vacuum Pump, Double Air Separation,fans, control system, Air Filter
2. When vacuum pump start working , Air Separation start working also, in order to drain and exhaust directly
3. Screen show working station
Struction and working theroy
1. Inside major structure--Vacuum Pump, Double Air Separation,fans, control system, Air Filter
2. When vacuum pump start working , Air Separation start working also, in order to drain and exhaust directly
3. Screen show working station
5.2 Installation conditions
Indoor: solid and flat ground without any shake.
Clean environment with good ventilation and without direct sunlight and harmful, corrosive gas.
It is recommended to provide a ventilator and some air outlets with a diameter of 38mm in the room where this device is installed, to discharge the exhaust gas and guarantee the fresh indoor air.
In addition, an underground sewage pipe with an inner diameter of 25mm should be equipped in the room to guarantee the removal of liquid waste from the suction system.
This device should be installed near the power socket with reliable grounding.
Keep this device far away from heat. No inflammable materials near it. Enough space should be provided near the suction system for ventilation.
No direct or indirect contact with damp objects is allowed in the installation site.
5.3 Requirements for normal operating environments
Ambient environment: +5℃ -- +40℃
Relative humidity: ≤70%
Atmosphere pressure: 700Hpa ~ 1060Hpa
5.4 Pipeline/signal line connection
A Connection of air intake pipe of the suction system
Please first connect the provided inlet pipe with the water inlet of the suction system:
Operate with your both hands: one holds the water inlet of the suction system and the other holds the inlet pipe and then rotate it inward.
Then, fasten it with the provided metal hose clamp
Connect the other end of the inlet pipe to the strong and weak exhaust pipes (or the preset underground main) on the dental chair
B Connection of exhaust pipe
Please connect one end of the provided exhaust pipe with the air outlet on the back of the suction system:
Hold one end of the exhaust pipe and then rotate it inward to connect to the air outlet.
Fasten it with the provided metal hose clamp.
Then, connect the other end of the exhaust pipe to the exhaust hole (or the preset exhaust manifold) on the wall.
Note: the exhaust pipe temperature will rise obviously when the suction system is working, so please use a metal pipe or heat resistant tube that could withstand an 80℃ high temperature as the exhaust extension tube.
C Connection of drainage pipe
Please connect one end of the provided drainage pipe with the water outlet on the back of the suction system:
Hold one end of the drainage pipe and then rotate it inward to connect to the water outlet.
Fasten it with the provided metal hose clamp.
Then, connect the other end of the drainage pipe to the sewer pipe.
Note: the drainage pipe in clinic cannot be installed higher than the water outlet.
D Connection of the suction system and the signal line of dental chair
The signal line used to connect this device to the dental chair is a multi-conductor line located near the power wire on the back of this device.
Connect the signal line with the heavy-current and light-current control system switch of the dental chair.
.
5.5 Use after connection
Power on, and turn on the green rocker switch at the lower right-hand corner on the device front, the power indicator light and the monitor will be on and the suction system enters a standby mode.
After confirming the signal switch on the back of the suction system
remove the strong or weak suction pipe on the dental chair and then the suction machine begins to work