Product Specification
Condition | New |
Brand | Tempo |
Warranty | YES |
Voltage (V) | 230 |
Product Descriptions
Tempo Laboratory Hot Air Oven | Humidity Oven Temp 5°C to 60°C
“TEMPO” make humidity ovens are manufactured and design for electrical & electronic industries for testing components and equipment as per ISO 9000:2015 Testing cement, batteries, cigarettes, or any other material demanding controlled temperature & humidity. It is also used in industries like pharmaceutical for accelerating ageing, food & beverage, packing, department of botany, cytology, plant & animal breeding.
- Temperature range – 5°C above ambient to 60°C
- Temperature accuracy – ±0.5°C
- Temperature uniformity – ±1°C
- Humidity Range – 5% Rh above ambient to 90% Rh
- Humidity Accuracy – ±2%Rh
- Humidity Uniformity – ±3%Rh
Construction
“TEMPO” make humidity oven is double walled constructed with backside as a triple wall. The 3” thick PUF insulation ensures a stable temperature and humidity & also a reduction in energy consumption. The motorized blower is at the back side of the chamber to develops a unique air flow system which ensures maximum uniform temperature & humidity condition inside the chamber. The unique air flow assures quick recovery after door openings.
Technical Specification Controlling System
The temperature is controlled by means of imported microprocessors based auto tune PID controller with CE mark & dual display of set value & process value. The humidity is controlled by means of imported microprocessors based auto tune PID controller with CE mark & dual display of set value & process value with direct %RH.
Air Circulation
The motorized blower at the backside of the chamber develops a unique air ow system which ensures maximum uniform temperature and humidity condition inside the chamber. The unique airflow assures quick recovery after the doors are opened.
Heating System
High-quality U shape SS tubular heaters are used for better heating.
Humidity System
Steam injection system (non-condensing type) & reservoir tank with water level arrangement with indicator.
Features
The full view observation glass/acrylic door a gasket is provided to observe sample inside the chamber. The interior working area is illuminated.
The specially designed adjustable stainless steel rod trays ensure uniform temperature distribution.
Safety Features
Additional safety thermostat to cut of dry heater supply in case temperature of overshoot. Float switch to cut off the wet heater supply in case of low water level. Audio & visual alarm are provided.
Models & Optional Features
Models | Capacity (litres) | Internal Dimensions H × W × D (cm) | Trays |
TI – 701 S / G | 100 | 50 × 50 × 40 | 2 |
TI – 702 S / G | 170 | 70 × 60 × 40 | 2 |
TI – 703 S / G | 200 | 60 × 60 × 60 | 2 |
TI – 704 S / G | 285 | 80 × 60 × 60 | 2 |
TI – 705 S / G | 325 | 90 × 60 × 60 | 3 |
Optional Features
- CFR 21 part 11 compliance software for data logger
- Data logger
- Data scanner complete with sensor
- SMS mobile alert system
- Door access security system
- Safety digital temperature controller
- Timer
Standard Models
Inside SS 304 with mirror polish & outside mild steel powder coated.
GMP Models
Inside SS 316 with mirror polish & outside stainless steel 304 with the dull buff finish.